Title
Intercomparison of radiation protection instrumentation in a pulsed neutron field
Contributors
F Trompier (Institute for Radiological Protection and Nuclear Safety)
,
M Volnhals (Helmholtz Zentrum München)
,
M Wielunski (Helmholtz Zentrum München)
,
M Silari (CERN)
,
J Röhrich (Helmholtz-Zentrum Berlin für Materialien und Energie)
,
S Mayer (Paul Scherrer Institut )
,
K Ott (Helmholtz-Zentrum Berlin)
,
M Luszik-Bhadra (Physikalisch-Technische Bundesanstalt (PTB))
,
N Golnik (Warsaw University of Technology)
,
E Hohmann (Paul Scherrer Institut )
,
A Leuschner (Deutsches Elektronen-Synchrotron DESY)
,
M Caresana (Politecnico di Milano)
,
A Denker (Helmholtz-Zentrum Berlin für Materialien und Energie)
,
A Esposito (IFNF-LNF)
,
M Ferrarini (CNAO)
,
GP Manessi (CERN and University of Liverpool)
Abstract
In the framework of the EURADOS working group 11, an intercomparison of active neutron survey meters was performed in a pulsed neutron ?eld (PNF). The aim of the exercise was to evaluate the performances of various neutron instruments, including commercially available rem-counters, personal dosemeters and instrument prototypes. The measurements took place at the cyclotron of the Helmholtz-Zentrum Berlin für Materialien und Energie GmbH. The cyclotron is routinely used for proton therapy of ocular tumours, but an experimental area is also available. For the therapy the machine accelerates protons to 68 MeV. The interaction of the proton beam with a thick tungsten target produces a neutron ?eld with energy up to about 60 MeV. One interesting feature of the cyclotron is that the beam can be delivered in bursts, with the possibility to modify in a simple and ?exible way the burst length and the ion current. Through this possibility one can obtain radiation bursts of variable duration and intensity. All instruments were placed in a reference position and irradiated with neutrons delivered in bursts of different intensity. The analysis of the instrument response as a function of the burst charge (the total electric charge of the protons in the burst shot onto the tungsten target) permitted to assess for each device the dose underestimation due to the time structure of the radiation ?eld. The personal neutron dosemeters were exposed on a standard PMMA slab phantom and the response linearity was evaluated.
Organisation
CI
,
STFC
Keywords
Proton
,
detection Pulsed
,
beam Neutron
,
Neutron survey meter
Funding Information
Related Research Object(s):
Licence Information:
Language
English (EN)
Type
Details
URI(s)
Local file(s)
Year
Journal Article
Nucl Instrum Meth A
737 (2014): 203-213.
doi:10.1016/j.nima.2013.11.073
2014
Showing record 1 of 1
Recent Additions
Browse Organisations
Browse Journals/Series
Login to add & manage publications and access information for OA publishing
Username:
Password:
Useful Links
Chadwick & RAL Libraries
SHERPA FACT
SHERPA RoMEO
SHERPA JULIET
Journal Checker Tool
Google Scholar