ePubs

The open archive for STFC research publications

Full Record Details

Persistent URL http://purl.org/net/epubs/work/53761
Record Status Checked
Record Id 53761
Title The performance of photemission beamline 6.1
Contributors
Abstract The performance of beamline 6.1 has been characterised over several commissioning periods and the results are presented here. The monochromator covers the energy range 20 - 280 eV with an energy resolution of 0.1% of the photon energy. It has been established that a 1200 l/mm blazed grating yields higher flux with lower second order content than a 1200 l/mm lamellar grating. Three electrostatic electron energy analysers are available for use; the fixed MCD HA100 with a resolution of 1.5% of the pass energy; the fixed CMA with a resolution of 2% of the pass energy; and the HA50 on a two axis goniometer with a resolution of 1% of the pass energy. Details are also given of improvements made to the monochromator.
Organisation SRS , MPW6.1 , SERC
Keywords Physics
Funding Information
Related Research Object(s):
Licence Information:
Language English (EN)
Type Details URI(s) Local file(s) Year
Report DL Technical Reports DL-SCI-TM94E. 1993. DL-SCI-TM94E.pdf 1993